Indexed metadata

Accurate evaluation of self-heterodyne laser linewidth measurements using Wiener filters

Markus Kantner, Lutz Mertenskötter

Source record

Source: Crossref

Published: Apr 27, 2023

DOI: 10.1364/oe.485866

Open original source ↗

Source abstract

Self-heterodyne beat note measurements are widely used for the experimental characterization of the frequency noise power spectral density (FN–PSD) and the spectral linewidth of lasers. The measured data, however, must be corrected for the transfer function of the experimental setup in a post-processing routine. The standard approach disregards the detector noise and thereby induces reconstruction artifacts in the reconstructed FN–PSD. We introduce an improved post-processing routine based on a parametric Wiener filter that is free from reconstruction artifacts, provided a good estimate of the signal-to-noise ratio is supplied. Building on this potentially exact reconstruction, we develop a new method for intrinsic laser linewidth estimation that is aimed at deliberate suppression of unphysical reconstruction artifacts. Our method yields excellent results even in the presence of strong detector noise, where the intrinsic linewidth plateau is not even visible using the standard method. The approach is demonstrated for simulated time series from a stochastic laser model including 1/ f -type noise.

Evidence graph

No public relationships recorded yet.

Integrity note: This page is a factual metadata record created by deterministic ingestion. It is not a claim that the work moves a mathematical frontier or has been independently verified.